37
PROGRAMMED OPERATION METHOD
TO ADD SEGMENT
1.
To choose “Append seg”
・
On setup item choice screen, have “Append seg” display and press jog dial.
※
”Append seg” is not displayed when ramp time of last segment number indicates “End”,
sort time indicates “Hold” and the number of repeat times is “limitless”. It is not
displayed when key lock mode is “on” and program which is being edited is in
operation.
・
Segment number increases by one, while remaining segment number decreases by one.
Then you go back to setup item choice screen and ramp time of appended segment
number is displayed.
Append seg
P01-01
rest
10
(
Add segment
)
↓
Ramp time
►STEP
P01-02
rest
9
(
Choice screen
)
・
Contents which are initially set up with appended segment are as follows. Segment
which is carried out at the stage of program preparation has the same contents.
Item name
Initial
value
Remarks
Ramp time
Step
Setup temp
0
℃
Soak time
Hold
Wait function
Off
Wait function is of no effect since Soak time
indicates “Hold”.
Fan spinning
10
It is effective only when option is installed.
Damper opening
0%
It is effective only when option is installed.
Event 1
~
4
Off
It is effective only when option is installed.
Repeat start segment
No
Repeat start segment is of no effect since
Soak time indicates “Hold”.
The number of repeat
times
Infinity
The number of repeat times is of no effect
since repeat start segment is of no effect.
Append seg
P01-01 rest 10
℃
FIXED TEMP
AUTO START
AUTO STOP
PROGRAM
Содержание DF Series
Страница 2: ......
Страница 62: ...59 SCHEMATICS DF811 ...
Страница 63: ...60 DF1011 ...
Страница 64: ...61 DH811 ...
Страница 65: ...62 DH1011 ...
Страница 66: ...63 ...
Страница 70: ...67 ...
Страница 71: ......