Turb PLUS 2000
Operation
50
100193 Rev 0 ba77254e01 11/2020
Date
Time Reading
Units Alarm 1 Alarm 2
6/9/2017 17:05 0.02
NTU
Off
Off
6/9/2017 17:06 0.02
FNU
Off
Off
6/9/2017 17:07 0.022
FNU
Off
Off
The file structure of the logged data will be in Year/Month/Day. At midnight of each day the
current day’s file is closed and a new day’s file is opened.
NOTICE
For moisture tightness and humidity control, the supplied USB plug
cover must be kept installed firmly when not being used. Failure to
correctly install the plug will affect desiccant life.
7.4.3
Modbus Communication
Modbus protocol communication is operational on all models.
Patented Ultrasonic Cleaning 1
This feature is used to continuously clean the flow through the cuvette. It is not intended to clean
cuvettes that are already dirty or replace manual cleaning entirely. See section 7.7 for instructions
on cuvette cleaning.
The system will increase the time between cleanings dramatically. Please note that the system
requires the use of a special cuvette which is included with the instrument. This cuvette must be
used for the system to operate correctly.
The system works by sending an ultrasonic frequency through spring connections into a piezo
transducer bonded to the bottom of a flow-through cuvette (see figure 10).
The system can detect that an incorrect cuvette is installed, an error has occurred in the
transducer or the transducer is not making contact with the spring connections. This error is
indicated by
CLN
being posted to the
lower
screen. Since this is an error condition, this may affect
the 4-20 mA and alarms depending on the setting of the ERLV (4-20 mA) and if an alarm is set up
to Error.
1
U.S. Patent No. 7,808,642