Turb PLUS 2000
Setup
, Configuration and Calibration
36
100193 Rev 0 ba77254e01 11/2020
To set the 4mA Levels:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
4MA
is displayed.
3. Adjust setting using the
and
buttons.
4. Press the
button when adjustments are complete to save this setting and move on to the 20mA
adjustment.
20mA Adjustment
This menu operates similarly to the previous menu. This menu outputs a constant 20 mA while
allowing for a small amount of adjustment. The adjustment limits are ± 1000 counts or about ± 1
mA.
This setting will be slightly different on each instrument as each Turb PLUS 2000 will be factory
set to 20.00mA.
1.
To set the 20mA Levels:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
20MA
is displayed.
3. Adjust setting using the
and
buttons.
4. Press the
button to accept the setting and move to the next menu or the
button to exit and return
to
AUTO
.