Setup
, Configuration and Calibration
Turb PLUS 2000
100193 Rev 0 ba77254e01 11/2020
27
In case of an error in the Turb PLUS 2000, the 4-20 mA reading can be used to indicate a
problem by sending the current to either
4.00
mA,
2.00
mA.
0.00
mA, or
OFF
.
The factory default setting is
OFF
and the 4-20mA is unaffected by any error condition.
To change the Error Level:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
ERLV
is displayed.
3. Select the desired
ERLV
by using the
and
buttons then press the
button to accept the setting
and move to the next menu or the
button to exit and return to
AUTO
.
6.2.2
Baud Rate - Selecting the RS–485 Baud Rate
The baud rate can be selected for the operation of the I/O port. Available baud rates are: 1200,
2400, 4800, 9600, or 19200.
To change the Baud Rate:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
BAUD
is displayed.
3. Select the desired
BAUD
rate by pressing the
or
buttons to change the displayed baud rate.
4. Press the
button to accept the setting and move to the next menu or the
button to exit and return
to
AUTO
.