Turb PLUS 2000
Setup
, Configuration and Calibration
34
100193 Rev 0 ba77254e01 11/2020
6.2.10
Display Units - Setting Displayed Units
The most common unit is
NTU
(Nephelometric Turbidity Units) however the instrument can
display in
FNU
(Formazine Nephelometric Units). All instruments are shipped from the factory set
in NTU mode.
To set/change the units:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until UNIT is displayed.
3. Use the
or
buttons to select the setting
NTU
or
FNU
.
4. Press the
button to accept it and move to the next menu or the
button to exit and return to
AUTO
.
6.2.11
Ultrasonic Cleaning - Cleaning Function
This allows for a selection menu to turn off the ultrasonic cleaning function if desired. The default
mode is
On
.
To set the ultrasonic cleaning to On/Off:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
CLN
is displayed.
3. Select
On
or
OFF
using the
and
buttons.
4. Press the
button to accept I the setting and move to the next menu or the
button to exit and return
to
AUTO
.