Turb PLUS 2000
Operation
48
100193 Rev 0 ba77254e01 11/2020
5.
To speed up the recognition, of the new desiccant, by the instrument, it may be necessary to reset the
instrument by disconnecting the sensor interconnect cable for 2 seconds and then reconnecting it.
NOTICE
Once the bag is opened, install the desiccant pouch immediately to prevent
premature degradation of the desiccant.
Starting the System
The following steps describe how to measure the turbidity of a sample using this instrument:
1. Apply power to the instrument and allow the unit to warm up (typically 45 minutes – 1 hour on initial
commissioning).
2. Check the cuvette to make sure it is tight before starting the flow. The cuvette should be handled at the
top ½” to prevent fingerprints from interfering with readings.
3. After the initial warm-up period and once a continuous process stream is flowing through the
instrument, the instrument will display the measured turbidity level of the sample by displaying it on the
LCD screen. In addition, the equivalent signal is provided on the analog (4-20 mA) output, or the digital
output, depending on the options selected.
4. Initial settings like Date and Time (see section 6.2.17) must be set now.
Figure 8: Installing the Desiccant