Hardware Troubleshooting
MedWeld 5000 Technical Reference Manual
9 - 5
M-032170
Solving Typical Problems
This table describes some of the typical problems you may see in the
field. The recommended procedure to troubleshoot each problem is
provided.
Problem
Possible Cause
Possible Solution
Circuit breaker
trips on power-
up.
SHUNT TRIP output on the I/O is
becoming active.
If YES, check the following:
- Primary transformer cables grounded?
- SCR breaking over at low potential, or
shorted?
Inverter is faulty.
Replace inverter assembly.
Circuit breaker is faulty.
Replace circuit breaker.
Circuit breaker
trips while
welding
The magnetic trip on the circuit
breaker is set too low. (This causes
the control to trip during the weld time
of a schedule.)
Adjust the circuit breaker’s magnetic trip
threshold.
SHUNT TRIP output on the I/O
module is active when the SCR is not
shorted.
Faulty firing card or I/O module. Replace
as necessary.
Intermittent ground.
Replace circuit breaker.