Hardware Troubleshooting
9 - 4
MedWeld 5000 Technical Reference Manual
M-032170
Control and DEP-
100S cannot
communicate
(DEP and MedLAN
LEDs may be off or
blinking)
Bad connection between
timer unit and DEP-100S.
Check connection. Verify that it is
secure.
Check for bad/faulty connector and/or
cable.
Cycle power.
If networking, no network
power supply or no external
120 VAC source con-
nected to the power supply.
Make sure you have a network power
supply and that external 120 VAC is
connected.
Faulty DEP-100S, WTC
timer or Network Power
Pack.
Replace as necessary.
Incorrect network address
for weld control.
Verify the welder ID and network
address of the device.
Communication over
the MedLAN channel
is inoperable.
(MedLAN LEDs may
be off or blinking)
Improper or loose MedLAN
wiring.
Check for secure and correct wiring.
Two or more controls have
the same network address
number.
Check MedLAN address of each
control. Each control must have a
UNIQUE address.
Faulty WTC timer, faulty
DEP-100S or faulty Net-
work Power Pack.
Replace as necessary.
Symptom
Probable Cause
Corrective Action
Содержание MedWeld 5000
Страница 2: ...MedWeld 5000 Technical Reference Manual Modified 1 31 06 M 032170...
Страница 6: ...MedWeld 5000 Technical Reference Manual Modified 1 31 06...
Страница 38: ...Installing the MedWeld 5000 2 12 MedWeld 5000 Technical Reference Manual M 032170...
Страница 82: ...Advanced Software Features 5 12 MedWeld 5000 Technical Reference Manual M 032170...
Страница 120: ...Hardware Troubleshooting 9 10 MedWeld 5000 Technical Reference Manual M 032170...