TABLE 11: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The CS CONTROL SYSTEM is intended for use in the electromagnetic environment specified below.
The customer or the end user of the CS CONTROL SYSTEM should assure that it is used in such an environment.
Immunity Test
EN/IEC 60601
Test Level
Compliance Level
Intended
Electromagnetic
Environment
Electromagnetic
Discharge (ESD)
EN/IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidi-
ty should be at least 30%.
Electrical fast transient/
burst
EN/IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/output
lines
± 2kV for power supply
lines
± 1kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
EN/IEC 61000-4-5
± 1kV differential mode
(line-line)
± 2kV common mode
(line-earth)
± 1kV differential mode
(line-line)
± 2kV common mode
(line-earth)
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply
input lines
EN/IEC 61000-4-11
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 seconds
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 seconds
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the CS
CONTROL SYSTEM requires contin-
ued operation during power mains in-
terruptions, it is recommended that the
CS CONTROL SYSTEM be powered
from an uninterruptible power supply
or a battery.
Power frequency
(50/60Hz) magnetic field
EN/IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
Note UT is the a.c. mains voltage prior to application of the test level.