LEVEL
•
Thermistor Status:
Gives the user the status for the 8-thermistors on the level sensor
assembly, the freezeguard sensor and the lid switch.
•
Sensor Positions:
Allows the user to set the positions for low level alarm, start fill, stop fill, high
level alarm and the sensor offset.
•
Valve Open Duration:
Allows the user to set the fixed times for valve open and valve closed
durations.
•
Sensor Error Mute:
Allows the user to mute the audible alarm for a sensor error.
•
Inch/MM:
Allows the user to choose units of measurement for level.
SYSTEM
•
Date/Time:
•
Set Date/Time:
Allows the user to set the date
and the time. Use the up/down arrows to
change settings. Use the left/right arrows to
move between the date and time fields.
•
Date Format:
Allows the user to change
between U.S. and International format.
•
Date & Time Sync:
This feature can be used to
automatically synchronize the date and time
between the CS200 Control and the ciSmart
Com Communications Hub.
•
Lid Functions
•
Lid Switch:
Allows the user to enable/disable the lid switch feature.
•
Quickchill Timer:
Allows the user to set the amount of time that the solenoid valve will stay open
after the lid is closed.
•
Manual Defog Timer:
Allows the user to set the amount of time that the solenoid valve stays open
when the Fill/Defog button is pressed and the LN
2
level is within normal range.
•
Auto Defog Timer:
Allows the user to set the amount of time that the solenoid valve stays open
when the lid is opened and LN
2
level is within the normal range.
•
User Access
•
User Access On/Off: Allows the user to enable/disable this feature.
•
User ID’s: Allows the user to set user identification for freezer access.
•
User ID On/Off: Allows the user to turn off user identification.
•
Delete user ID: Allows the user to delete user ID’s.