CS200
Control Systems
28
Figure 6.1 Menu System - Change Settings
Controller
Operation
MENU
CHANGE SETTINGS
TEMPERATURE
THERMOCOUPLE SELECT
T/C 1
On/Off
T/C 2
On/Off
CALIBRATE
CALIBRATE THERMOCOUPLE 1
PUT T/C 1 IN ICE WATER (0C) PUT
T/C 1 IN LN2 (-196C) DONE!
CALIBRATE THERMOCOUPLE 2
PUT T/C 2 IN ICE WATER (0C) PUT
T/C 2 IN LN2 (-196C) DONE!
RESTORE CALIBRATION
ABORT/LOAD
TEST TEMPERATURE SYSTEM
CHECK THERMOCOUPLES
T/C 1 Ok/Off
T/C 2
Ok/Off
TEST TEMP ALARMS
Off- High Temp 1!- High Temp2!- Low
Temp1!- Low Temp2!
CONTROL BY TEMPERATURE
TEMP CONTROL ON/OFF
On/Off
TEMP CONTROL RANGE
-170:-105 C -165:-100 C
TEMPERATURE ALARMS
HIGH TEMPERATURE ALARM
THERMOCOUPLE 1
-190:0
THERMOCOUPLE 2
-190:0
LOW TEMPERATURE ALARM
THERMOCOUPLE 1
-196:0 On/Off
THERMOCOUPLE 2
-196:0 On/Off
TIME WITH NO TEMP ALARM
TEMP1: 0
TEMP2:0
TEMPERATURE UNIT
CELCIUS
FAHRENHEIT KELVIN
RANKINE
LEVEL
THERMISTOR STATUS
123456789
LID SWITCH
LLLGGGGGG
Open/Closed
Thermistor Status
0:
1:
2:
3:
4:
5:
6:
7:
8:
SENSOR POSITIONS
LLA START STOP HLA
02 03 06 07
SENSOR OFFSET 0:60 INCH
VALVE OPEN DURATION
OPEN 0:120 MIN CLOSE 0:60 MIN
SENSOR ERROR MUTE
On/Off
INCH/CM
INCH/cm/mm
SYSTEM
DATE/TIME
SET DATE/TIME
00 - 00 - 00
00: 00
DATE FORMAT
U.S. INTERNATIONAL
US:
MM/DD/YY INTL: DD/MM/YY
DATE & TIME SYNC
On/Off
LID FUNCTIONS
LID SWITCH
On/Off
QUICKCHILL TIMER
0:300 seconds
MANUAL DEFOG TIMER
0:300 seconds
AUTO DEFOG TIMER
0:300 seconds
USER ACCESS
USER ACCESS ON/OFF
On/Off
USER IDs
USER1
USER2
USER3
USER ID ON/OFF
On/Off
DELETE USER ID
USER1
USER2
USER3
LID SAFETY INTERLOCK
On/Off
SYSTEM ALARMS
TEST ALARMS
AUDIBLE
On/Off
VISUAL
On/Off
REMOTE
On/Off
LN2 SUPPLY ALARM DELAY
30-45-60-75-90 MIN
2-3-4
HOURS
REMOTE ALARM DELAY
IMMEDIATE 30-60-90 MIN
2-3-4-6 HOURS
LID OPEN ALARM DELAY
1:20 MIN
VALVE STUCK ALARM
On/Off
AUDIBLE REPEAT DELAY
15-30-60-120-240-480 MIN
1
DAY NEVER
USAGE ALARM
On/Off CLEAR ALARM
DISPLAY
DISPLAY BRIGHTNESS
1-15
FREEZER ID
FREEZERID
FILL SEQUENCE NUMBER
0-100
LOCAL/NETWORK
(0 = NO SEQUENCING)
SCHEDULED EVENTS
SCHEDULED FILLS
SET TIME
SET DAY
SET TIME
Daily-
Off-Day 00:00-24:00
NEXT FILL
00-00-00
00:00
MAINTENANCE REMINDER
ACKNOWLEDGE REMINDER
YES
NEXT REMINDER
00-00-00
00:00
LN2 SUPPLY REMINDER
SET REMINDER
SET DAY
SET TIME
Daily-
Off-Day 00:00-24:00
ACKNOWLEDGE REMINDER
YES
NEXT REMINDER
00-00-00
00:00
CHECK LEVEL REMINDER
SET REMINDER
SET DAY
SET TIME
Daily-
Off-Day 00:00-24:00
ACKNOWLEDGE REMINDER
Enter Level Measured
0:1250
(in .5" increments)
NEXT REMINDER
00-00-00
00:00
SECURITY
SETTINGS PASSWORD
0 0 0 0
POWER PASSWORD
0 0 0 0
CHANGE REMOTE PASSWORD
0 0 0 0
CONTACT SUPPORT
LN2 SUPPLIER
CONTACT INFO
DISTRIBUTOR
CONTACT INFO
WORTHINGTON
CONTACT INFO
PACER DIGITAL
CONTACT INFO
SELECT LANGUAGE
ENGLISH-GERMAN
SYSTEM INFORMATION
SERIAL # 12345
VERSION 1. 03
FREEZER ID FRZ01
Options
DATA MANAGEMENT
DOWNLOAD LOGS
00-00-00 All Logs-Temp Logs- Alarm
Logs- Temp&Alarm Logs
USAGE
0.00 IN PER DAY
Thermistor Status
0: 1: 2:
3: 4: 5:
6: 7: 8:
30-45-60-75-90 MIN
2-3-4 HOURS
15-30-60-120-240-480 MIN
1 DAY NEVER
SET DAY
SET TIME
Daily-Off-Day 00:00-24:00
SET DAY
SET TIME
Daily-Off-Day 00:00-24:00
SET DAY
SET TIME
Daily-Off-Day 00:00-24:00
Enter Level Measured
0:1250 (in .5” increments)