![WIKA IPT-2 Series Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/wika/ipt-2-series/ipt-2-series_operating-instructions-manual_980233058.webp)
58
9 Supplement
WIKA Operating Instructions - Process pressure transmitter IPT-2x
Material cable gland/Seal insert
Cable diameter
5 … 9 mm
6 … 12 mm
7 … 12 mm
Brass nickel-plated/NBR
●
●
–
Stainless steel/NBR
–
–
●
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP68 (25 bar)
Connection cable transmitter - external housing, mechanical data
Ʋ Configuration
Wires, strain relief, breather capillaries, screen braiding,
metal foil, mantle
17)
Ʋ Standard length
5 m (16.40 ft)
Ʋ Max. length
180 m (590.5 ft)
Ʋ Min. bending radius at 25 °C/77 °F
25 mm (0.985 in)
Ʋ Diameter
approx. 8 mm (0.315 in)
Ʋ
Material
PE, PUR
Ʋ Colour
Black, blue
Connection cable transmitter - external housing, electrical data
Ʋ Wire cross-section
0.5 mm² (AWG 20)
Ʋ Wire resistance
0.037 Ω/m (0.012 Ω/ft)
Interface to the Master sensor
Data transmission
Digital (I²C-Bus)
Connection cable Slave - Master, mechanical data
Ʋ Configuration
Cores, strain relief, braided, metal foil, jacket
Ʋ Standard length
5 m (16.40 ft)
Ʋ Max. length
25 m (82.02 ft)
Ʋ Min. bending radius (at 25 °C/77 °F)
25 mm (0.985 in)
Ʋ Diameter
approx. 8 mm (0.315 in), approx. 6 mm (0.236 in)
Ʋ
Material
PE, PUR
Ʋ Colour
Black
Connection cable Slave - Master, electrical data
Ʋ Wire cross-section
0.34 mm² (AWG 22)
Ʋ Wire resistance
< 0.05 Ω/m (0.015 Ω/ft)
Voltage supply for the complete system through Master
Operating voltage
Ʋ
U
B min
12 V DC
Ʋ
U
B min
with lighting switched on
16 V DC
17)
Breather capillaries not with Ex-d version.