2. Under the Users section in Settings, edit the
current user and check “yes” for use Evident
Hub.
3. Login to Evident Hub using the email address
for your Evident account.
7. Maintenance
7.1 Cleaning the Handpiece
The entire body, cord and base of the scanner must be wiped down using a Federal
Environmental Protection Agency (EPA) approved disinfectant that is labeled and specified
for tuberculocidal/ mycobactericidal activity. Do not use disinfectant on the nozzle.
NOTE:
All components of the scanner (excluding the tips) must be
wiped down and not sprayed. Avoid getting any moisture, alcohol or
disinfectant inside the open scanner chamber.
Recommended and approved surface disinfectants:
• Birex Wipes: TB Claim= 10 minutes- ‘Phenolic (Dual) Water-Based’
• Prospray Wipes: TB Claim=10 minutes- ‘Phenolic (Dual) Water-Based’
• Cavicide Wipes: TB Claim=3 minutes- ‘Phenolics (Dual) Alcohol-Based’
• DisCide ULTRA Wipes: TB Claim-1 minute- ‘Phenolics (Dual) Alcohol-Based’
• Maxiwipe Germicidal Cloth: TB Claim=5 minutes- ‘Phenolics (Dual) Alcohol-Based’
• Ster 1 Plus: TB Claim=5 minutes- ‘Quaternary ammonium and Alcohol-Based’