VeriPulse IOS User Manual
53
Immunity Test
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
<5% UT (>95%
dip in UT) for
0,5 cycle 40%
UT (60% dip in
UT) for 5 cycles
70 % UT (30%
dip in UT) for
25 cycles
<5% UT (>95%
dip in UT) for
5 s
<5% UT (>95%
dip in UT) for
0,5 cycle 40%
UT (60% dip in
UT) for 5 cycles
70 % UT (30%
dip in UT) for
25 cycles
<5% UT (>95%
dip in UT) for
5 s
Mains power quality should be that
of a typical professional healthcare
facility environment.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of
a location in a typical professional
healthcare facility environment.
Radiated RF
IEC 61000-4-3
Table 9 in IEC-
60601-1-2 2014
Table 9 in IEC-
60601-1-2 2014
Portable and mobile RF
communications equipment should
be used no closer to any part of
the VeriPulse IOS system, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
where P is the maximum output
power rating of the transmitter
in watts (W) according to
the transmitter manufacturer
and d is the recommended
separation distance in metres
(m). Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Conducted RF
IEC 61000-4-6
3V 0.15-80MHz
6V in ISM
bands between
0.15 MHz and
80 MHz
80% AM at
1KHz
3V 0.15-80MHz
6V in ISM
bands between
0.15 MHz and
80 MHz
80% AM at
1KHz
NOTE: UT is the A.C. mains voltage prior to application of the test level.