VeriPulse IOS is intended for use in a professional healthcare setting with electromagnetic
environment specified below.
Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment
Guidance
RF Emissions CISPR 11
Group 1, Class A
VeriPulse IOS uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
Harmonic emissions IEC
61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions IEC
61000- 3-3
Complies
VeriPulse IOS is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Electromagnetic Immunity
Immunity Test
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge
± 8 kV /
Contact
± 8 kV /
Contact
Surface should be wood, concrete,
or ceramic tile. If floors are covered
(ESD) IEC
61000-4-2
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV /air
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV /air
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
± 2 kV for
power supply
lines ± 1 kV for
input/output
lines
± 2 kV for
power supply
lines ± 1 kV for
input/output
lines
Mains power quality should be that
of a typical professional healthcare
facility environment.
Surge IEC
61000- 4-5
± 0.5 kV, ± 1 kV
line(s) to line(s)
± 0.5 kV, ± 1 kV,
± 2 kV line(s) to
earth
± 0.5 kV, ± 1 kV
line(s) to line(s)
± 0.5 kV, ± 1 kV,
± 2 kV line(s) to
earth
Mains power quality should be that
of a typical professional healthcare
facility environment.