VeriPulse IOS User Manual
46
7.2 Cleaning and Sterilization of Tip
NOTE:
The included tips must be autoclaved prior to use as they do
not come pre-sterilized.
NOTE:
Ensure that the surface of the mirror does not show residues,
smudges, scratches, or any damage, as this would affect the
performance of the device.
Step-by-Step Procedure
Step 1:
Clean the tip with soapy water, ensuring that the mirror
is clean and free of smudges, stains, or any residue. Avoid using
abrasive cloth materials as this will scratch the mirror.
Step 2:
After drying the tip exterior, carefully dry the interior and
mirror with lint free wipes ensuring you do not scratch the surface.
The mirror should be free of any noticeable debris or water spots.
Step 3:
Insert and seal the tip into a sterilization pouch. Make sure
the seal is airtight. Each tip should be packaged individually.
Step 4:
Sterilize the wrapped tip in a steam autoclave using the
following parameters:
• 132°C (270°F) at 4 minutes, or
• 134°C (273°F) at 4 minutes
• 121°C (250°F) at 45 minutes
Step 5:
Ensure the dry cycle is complete prior to removing the tip
from the autoclave. If the pouch is damp with moisture, proper
sterilization cannot be guaranteed.7.3 Disposal
The VeriPulse IOS is an electrical device with electronical
components inside and should be disposed of in accordance with
local environmental laws and regulations.
WARNING:
Always autoclave the tip wrapped in a sealed sterilization
pouch; failure to do so will result in permanent stains on the mirror.