Electromagnetic immunity
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment
and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or user of the device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the device
as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated max. output
power of
transmitter (W)
150 kHz to 80 MHz
outside ISM bands
150 kHz to 80 MHz
in ISM bands
80 MHz to 800 MHz 800 MHz to 2.7 GHz
0.01
0.12
0.20
0.12
0.23
0.1
0.37
0.63
0.38
0.73
1
1.17
2.00
1.20
2.30
10
3.69
6.32
3.79
7.27
100
11.67
20.00
12.00
23.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
Test specifications for enclosure port immunity to RF wireless communications equipment
Test frequency
(MHz)
Band
a
MHz
Service
a
Modulation
b
Maximum
power (W)
Distance
(m)
Immunity
test level
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
b
18 Hz
1.8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM
c
±5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE band 13, 17 Pulse
modulation
b
0.2
0.3
9
146 General compliance and standards
Welch Allyn RetinaVue™ 700 Imager
Содержание RetinaVue 700
Страница 1: ...Welch Allyn RetinaVue 700 Imager Directions for use Software version 1 XX...
Страница 21: ...Directions for use First time set up of the RetinaVue 700 Imager 17...
Страница 26: ...22 First time set up of the RetinaVue 700 Imager Welch Allyn RetinaVue 700 Imager...
Страница 102: ...98 Using the RetinaVue 700 Imager Welch Allyn RetinaVue 700 Imager...
Страница 120: ...116 Controls Welch Allyn RetinaVue 700 Imager...
Страница 140: ...136 Device radio Welch Allyn RetinaVue 700 Imager...
Страница 152: ...148 General compliance and standards Welch Allyn RetinaVue 700 Imager...
Страница 156: ...152 RetinaVue 700 Imager Accessories Welch Allyn RetinaVue 700 Imager...
Страница 158: ...Material No 729216...