Electromagnetic immunity
±0.5 kV, ±1 kV, ±2 kV
Line-to-ground
±2 kV
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
0 % U
T
; 0.5 cycle
0 % U
T
; 0.5 cycle
Mains power quality should be
that of a typical professional
healthcare facility environment. If
the user of the device requires
continued operation during power
mains interruptions, it is
recommended that the device be
powered from an uninterruptible
power supply or a battery.
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25/30 cycles
Single phase: at 0°
70 % U
T
; 25/30 cycles
0 % U
T
; 250/300 cycle 0 %U
T
; 250/300 cycle
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
professional healthcare facility
environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.
144 General compliance and standards
Welch Allyn RetinaVue™ 700 Imager
Содержание RetinaVue 700
Страница 1: ...Welch Allyn RetinaVue 700 Imager Directions for use Software version 1 XX...
Страница 21: ...Directions for use First time set up of the RetinaVue 700 Imager 17...
Страница 26: ...22 First time set up of the RetinaVue 700 Imager Welch Allyn RetinaVue 700 Imager...
Страница 102: ...98 Using the RetinaVue 700 Imager Welch Allyn RetinaVue 700 Imager...
Страница 120: ...116 Controls Welch Allyn RetinaVue 700 Imager...
Страница 140: ...136 Device radio Welch Allyn RetinaVue 700 Imager...
Страница 152: ...148 General compliance and standards Welch Allyn RetinaVue 700 Imager...
Страница 156: ...152 RetinaVue 700 Imager Accessories Welch Allyn RetinaVue 700 Imager...
Страница 158: ...Material No 729216...