Problem
Solution
SSID Your Network
Access Point MAC B2:DE:31:OC:
32:B2
RSSI -70
Message:
Failure to connect to the portal. Unable
to delete patient information
.
Follow all steps for connecting to
RetinaVue Network Server.
Clinic network does not appear in the Wi-Fi list.
Ensure that your network meets the
supported credentials type.
Check with your network administrator
and ensure the network you are
attempting to connect with has one of
the following security types:
Enterprise WPA2-PSK, PEAP-
MSCHAPv2, EAP-TLS with certificates
Hidden SSID networks
Manually enter the network name.
Touch
Settings
.
Touch
Wi-Fi
.
Scroll to the bottom of the list. Touch
Other
and launch the keyboard to
enter the network name, followed by
password, and then select security
type.
Directions for use
Troubleshooting 123