Configuration software
47
Feature
Description
Unit temp
sample period
The time between samples of the internal temperature sensor.
The module temperature (Cold junction temperature) is sampled
every time the thermocouple input is sampled. You will only need
to set this field if you want to sample the temperature more often
than you sample the thermocouple, including when you are not
using a thermocouple.
Unit temp warm
up time
The amount of time the analog circuits are turned on before
sampling the internal temperature sensor.
You should always set this value set to the minimum value (0.08
seconds).
Supply voltage
sample period
Time between samples of the internal supply voltage.
Ain1 debounce
The debounce time is the product of the number of samples and
the sample period. The signal must differ from the previous value
by the sensitivity value for this duration to register a ‘change of
state’.
This means that changing the analog sample time will also change
the debounce time. For analog signals, a value of 0 or 1 generates
the quickest response; 255 generates the slowest response.
Ain1 sample
period
Sets the period between samples of the thermocouple input.
A smaller value will cause mappings using this input to be
triggered more frequently (i.e. quicker response).
A larger value will cause mappings using this input to be triggered
less frequently (less power consumption).
Ain1 warm up
time
The amount of time the analog circuits are powered up before
sampling the analog input.
Din1 Debounce
The amount of time the first digital input must stay in a changed
state before it registers as a change of state.
Din2 Debounce
The amount of time the second digital input must stay in a
changed state before it registers as a change of state.