5-24
Maintenance Procedures
1. Retrieve the combined 2.5-mm Allen wrench and cone extraction tool
from its storage location on the source adaptor housing.
2. On the combined 2.5-mm Allen wrench and cone extraction tool, slide
the collar to the end of the tool.
3. Insert the collar in the sample cone.
Collar