5-42
Maintenance Procedures
12. Holding the PEEK terminal block gently, use the needle-nose pliers to
gently grasp the heat-shrink tubing on the heater cartridge assembly
and carefully slide it and the PEEK terminal block out of the ion block.
13. Using the O-ring removal kit, carefully remove the cover seal from the
ion block (see
).
Heat-shrink tubing
Heater cartridge assembly
PEEK terminal block
Cover seal
Cone gas O-ring