Replacing the IntelliStart Fluidics tubing 5-109
Required materials
•
Chemical-resistant, powder-free gloves
•
SYNAPT G2 fluidics tubing and fitting kit
Tip:
This kit contains components for both the sample and LockSpray
system plumbing.
To plumb the analyte system
1. Open the access doors to the IntelliStart Fluidics system.
2. Using the stainless steel tubing, connect the sample pump to port 7 of
the sample selector valve.
Warning:
The IntelliStart Fluidics tubing and fittings can be
contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while working on the
system.
Caution:
The steel tubing requires high-pressure seals at each
end. Ensure that you tighten the fittings appropriately.
A
B
C
A
B
C
A
B
C
Waters
A
B
C
Sample selector valve
Lock-spray pump