Connecting the NanoLockSpray source camera
B-23
Connecting the NanoLockSpray source camera
The NanoLockSpray™ source camera is connected to the MassLynx™
workstation by means of a USB converter.
To connect the camera to the MassLynx workstation
1. Connect the coaxial cable between the video output connector on the
instrument’s rear panel and the interface box.
2. Connect the USB cable between the interface box and any free USB
socket on the MassLynx workstation.
Installing the camera driver software
The MassLynx workstation will be supplied with the camera driver software
ready installed. However, if you reinstall MassLynx, you will need to reinstall
the camera driver software using the Found New Hardware Wizard for the
USB Converter.
Installing the camera driver software
1. In the Found New Hardware Wizard dialog, click No, not at this time.
2. Click Next.
3. Click Install from a list or specific location (Advanced).
4. Click Next.
5. Select Search for the best driver in these locations.
6. Select Include this location in the search: and browse to
C:\MassLynx\USBCameraDriver.
External Connections 1
10
1
EPC Com Port
Video Output
Video output connector
MassLynx
workstation
USB converter