5-108 Maintenance Procedures
Plumbing the IntelliStart Fluidics sample delivery system
This section explains how to plumb the sample delivery system for standard
flow applications. For low flow, nanoACQUITY UPLC applications, the
column outlet passes directly to the nanoflow sprayer, bypassing the
IntelliStart Fluidics system.
IntelliStart Fluidics sample delivery tubing schematic
All tubing is PEEK, with the exception of the connection between the pump
and sample selector valve, which is stainless steel.
Tubing dimensions—sample delivery system
Valve / Port
Connection
ID (inch)
OD
(inch)
Color
Length
(mm)
Sample / 1
Waste reservoir
0.040
1/16
Natural
1000
Sample / 2
Wash reservoir
0.020
1/16
Orange
1000
Sample / 3
Divert valve
0.005
1/16
Red
200
Sample / 4
Vial A
0.020
1/16
Orange
680
Sample / 5
Vial B
0.020
1/16
Orange
680
Sample / 6
Vial C
0.020
1/16
Orange
680
Sample / 7
Analyte pump
0.040
1/16
n/a
500
Divert / 4
Waste
0.040
1/16
Natural
1000
A
B
C
2
3
4
1
1
6
5
4
3
7
2
From pump
From wash bottle
To probe
From UPLC
Waste
port
Tubing
guides
Sample selector valve
Diverter
valve
Содержание SYNAPT G2
Страница 18: ...xviii Table of Contents...
Страница 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Страница 66: ...3 20 Configuring the LockSpray Source...
Страница 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Страница 232: ...C 4 Materials of construction and compliant solvents...