Configuring for ESI mode
3-3
1. Prepare the instrument for working on the source (see
2. Remove the protective sleeve, if fitted, from the ESI probe tip.
3. Carefully slide the ESI probe into the hole in the probe adjuster
assembly, ensuring that the probe location dowel aligns with the
location hole in the probe adjuster assembly.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds,
handle the ESI probe with care.
TP03129
Location hole in the probe
adjuster assembly
ESI probe location dowel