Replacing the IntelliStart Fluidics tubing 5-101
Required materials
•
Chemical-resistant, powder-free gloves
•
SYNAPT G2 fluidics tubing and fitting kit
Tip:
This kit contains components for both the sample and LockSpray
system plumbing.
•
For the LockSpray reference probe connection
–
375 mm of 0.005-inch ID red PEEK
–
Long "finger tight" fitting
•
For the NanoLockSpray source reference probe connection
–
375 mm of 25-µL, fused-silica capillary (supplied in 1-m lengths)
–
Protective sleeve consisting of 0.015-inch ID PEEK tubing
–
F-130 "finger tight" nut
To plumb the LockSpray system
1. Open the access doors to the IntelliStart Fluidics system.
2. Using a PEEK nut, Super Flangeless™ ferrule, and stainless steel (SS)
ring, connect the blue, 1/16-inch, 300-mm, PEEK tubing from the
LockSpray pump to port 7 on the LockSpray selector valve, where you
use the long "finger tight" fitting.
Warning:
All tubing and fittings can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while working on the
system.