Replacing the IntelliStart Fluidics tubing
5-99
Removing the IntelliStart Fluidics tubing
This procedure explains how to remove the IntelliStart Fluidics tubing and
disconnect the probe tubing at the diverter valve or grounded union.
Required materials
Chemical-resistant, powder-free gloves
To remove the tubing
1. Open the access doors to the IntelliStart Fluidics system.
2. Unscrew and remove the PEEK thumbscrews and tubes from the
LockSpray selector, sample selector and diverter valves.
3. Unscrew and remove the PEEK thumbscrews and tubes from between
the flow sensor and the grounded union.
4. Close the access doors to the IntelliStart Fluidics system.
5. Dispose of the tubes and fittings in accordance with local environmental
regulations.
Plumbing the IntelliStart Fluidics LockSpray system
This section describes how to plumb the LockSpray system.
Requirement:
Finger tighten all PEEK fittings.
Warning:
All tubing and fittings can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while working on the
system.
Warning:
The IntelliStart Fluidics tubing and fittings can be
contaminated with biohazardous and/or toxic materials.
Dispose of them according to local environmental
regulations.
Содержание SYNAPT G2
Страница 18: ...xviii Table of Contents...
Страница 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Страница 66: ...3 20 Configuring the LockSpray Source...
Страница 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Страница 232: ...C 4 Materials of construction and compliant solvents...