34
vwr.com I Instruction manual VWR® U Water System
3.2.6.2 Install and Register Cartridges
VWR U system has a total of up to 7 packs, 1 or 2 RO packs, 1 AC Pack, 1 T Pack (Optional), 1 P
Pack, 1 H Pack, 1 U Pack. Each cartridge’s position is labelled on the system. Install a cartridge to
its corresponding position.
Cartridge installation is illustrated below.
a) Remove the right side panel from the system as shown below.
b)
Take out the cartridges from their packaging, respectively. Remove caps on the cartridge’s
inlet and outlet on the system.
c) Follow the instruction below to install the U pack to its corresponding position.
d) Wet the O-rings on the U Pack cartridge with pure water.
e) Turn the lever on the cartridge adapter to the right side at the unlock position as shown in
the photo.
P Pack
AC Pack
H Pack
U Pack
RO Pack