48
vwr.com I Instruction manual VWR® U Water System
4 DISPENSING WATER
Water can be dispensed from the dispenser handle either by pressing the dispense button on the
handle or from the main control console.
4.1 Dispensing RO Water
4.1.1 Press the dispense button on the dispenser to dispense RO water. Press the dispense
button again to stop dispensing. Or
4.1.2
Tap the left cylinder icon to decrease the volume
to 0 L. Tap the dispense button to dispense
RO water. Tap it again to stop dispensing.
4.1.3 Open the water faucet to dispense RO water directly from the pure water tank.
Note:
Dispensers are interlocked. Only one dispenser can be used at a time. Other dispensers will
display “In Use” on the dispenser.
4.2 Dispensing UP Water
4.2.1 Tap recirculation icon on the main console or on the dispenser to start
recirculation. When the resistivity reaches 18.2 MΩ.cm, press the dispense icon on the monitor
or on the dispenser to dispense UP water. Press the dispense icon again to stop dispensing.
4.3 Adjusting Dispense Rate
To change the dispense rate, tap the icon or on home screen or dispenser monitor.
4.4 Volumetric Dispensing
Setting up is the same on the dispenser or the
control console.
Tap recirculation icon on the home screen to
enter Recirculation mode until the resistivity
reaches 18.2 MΩ.cm.