27
vwr.com I Instruction manual VWR® U Water System
i)
Install the UV sanitization module.
Step 1 Open the box, and take out the three parts of UV sanitization module: sanitization
module controller, UV lamp, module body (including a cover, a quartz chamber and a bracket).
Step 2 Screw the tank cap to the tank top, and then screw off the hexagonal nut on the
module body.
Step 3 Push the signal cable with the UV lamp base and the sealing cap through the hole in
the middle of the controller. Then carefully insert the UV lamp completely into the chamber.
Hold the UV lamp and connect it to the UV lamp base. Screw the sealing cap to the UV lamp
mounting hole.
Step 4 Remove the nuts on the back of the controller. Then fix the controller to the support
with the nuts.
Step 5 Connect the signal cables to ASM port and OVER FLOW port on the system
respectively.
Nut
Sealing Cap
Tank Cap
Controller
Support
Controller
Tank Cap
Connect to System
Overflow Sensor
Overflow
Sensor
Sealing Cap
Module body
Tank Sanitization Module
Ballast and Cables
Connected to
System ASM
Tank UV
To UV
Lamp Base
Quartz housing