61
vwr.com I Instruction manual VWR® U Water System
5.4.6 Replacing the Final Filter
a) Remove the final filter
b) Install the filter into UP outlet firmly and make sure no leaking. Dispense UP water to degas.
Stop UP dispensing, installation finished.
c) If a final ultrafiltration cartridge is needed, install the device the same way as the final filter.
Note: Any change of type of final filter, need to reconfigure the final filter types
in the setting menu( Maintenance menu
→
Setting
→
Final Filter)!
5.4.7 Replacing the
pipe filters
a) Switch the system to Standby mode, Tap the dispense button
on the dispenser to dispense
water. Gas inside the system will be released out of the system. then remove the system right side
panel.
b) Turn the lever on the cartridge adapter to right to the unlock position, pull the cartridge
downwards to disengage the cartridge from its adapter. Remove all the cartridges.
c) Using a Phillips screwdriver, remove the 5 screws in the red circle on the white back panel.
d)
Find the pipe filters shown as the red circle in the left side picture below, remove the blue
buckles shown as the yellow circle in the right side picture below.
e) Remove the pipe filters and install a new one.
Note: Please confirm each filter needs to be placed in the right direction.