35
vwr.com I Instruction manual VWR® U Water System
Unlocked position Locked position
f)
Align the cartridge label facing out. Insert the cartridge’s up end first into the opening on the
cartridge adapter, then turn the lever handle to the left to lock the cartridge.
g) A pop-up dialog will be shown as below, tap
Install
to complete the installation.
Follow the same procedure to install other cartridges. Install the AC Pack, the P pack and the T
Pack
(Optional) cartridge to the left side of the system, the RO Pack(s) and the H Pack to the right
side.
Note!
The control console will display
“RFID Error!”, and the alarm icon
if a pack is installed
incorrectly. Reinstall it to the correct position.
h) Make sure all consumables have been registered and tap
Next
to enter Dispenser setup
Cartridge lock
lever
Lock Level