8.4.2 Verifying force measurement accuracy
Verification of force measurement accuracy requires a quiet environment as well as a rigid and stable work
surface.
Equipment required:
-
Force measurement accuracy verification kit
VOH 26.01001.SP11
-
Standard spindle
VOH 26.01001.SP04
Fitting procedure
PAViX screen
Comment(s)
PAViX
Remove the spindle and cleat in position under the
spindle.
To facilitate cleat removal, the spindle automatically
moves to a high position during initialisation.
Start by placing the standard spindle (SP.04 “EB”) on
the plate. Confirm with the right-hand side button.
Then place the calibration plate in place of the spindle,
resting on the standard spindle. Validate using the right-
hand side button
The PAViX automatically tares the force: do not
touch the calibration plate of the PAViX plate.