VOH SA
GU-04-06-01
PAViX - User Guide
Version 3.2
Created 10.2017
Page 43 / 48
8.4 Sensor inspection
8.4.1 Access to the Maintenance menu
The Maintenance mode is used to inspect distance and force sensors.
This is done by a validating the force or distance measurement by using a weight or a gauge block. Access to
the Maintenance mode is done using the following password:
7426
PAViX screen
Comment(s)
PAViX
The Maintenance menu is accessible via one of the
configuration pages: press and hold the PAViX left-hand
side button to enter the Setting mode.
Then, by rotating the barrel twice to the left, the
Maintenance menu will appear.
Confirm access using the right-hand side button.
The barrel allows the user to select the number to display,
the right-hand side button is used to confirm the choice.
The barrel is used to select the validation to perform:
-
Distance
: “Position[um]”
-
Force:
“Valid[N]”
Содержание PAViX