Validation procedure:
PAViX screen
Comment(s)
PAViX
…
1) Delicately place the required mass on the plate.
Confirm placement using the right-hand side
button.
2) Once the measurement is stable, confirm using
the right-hand side button.
These 2 operations must be performed with 1, 2 and 5kg
masses.
If the measures are correct, the screen displays “OK”.
Upon the first measure with an out-of-machine-
tolerance value, the operator is invited to recalibrate
the force.
(Refer to § 8.4.3 Force calibration, page 46)