PAViX - User Guide
Version 3.2
Created 10.2017
Page 46 / 48
8.4.3 Force calibration
Force calibration requires a quiet environment as well as a rigid and stable work surface.
Equipment required:
-
Calibration plate
VOH 26.01003.SP49
-
Set of standard weights for calibration
VOH 26.01003.SP50
-
Standard spindle
VOH 26.01001.SP04
Validation procedure:
PAViX screen
Comment(s)
PAViX
If force validation gives an out-of-machine-specification
result, you are prompted to calibrate the force
measurement system.
Remove the spindle and cleat in position under the
spindle.
Then place the calibration plate in place of the spindle,
resting on the standard spindle. Confirm with the right-
hand side button.
Once the measure is stable, the “check mark” button
appears.
It allows the operator to move to the next step.
Like on the image
below, with no weight.
Repeat the operations opposite for 1, 2, 5 and 10kg
masses.
Once calibration is complete, the error relating to the
extent of measurement is displayed (<= 2.0).
If the error on the extent of measurement is >2.0, it is not
possible to save the calibration.
3 consecutive calibrations can be carried out with a result
>2.0; after which, VOH SA must be contacted.