• Apnea alarm is possible to set to longer times and is an important
mean of surveillance to make sure that the patient gets ventilation
support regularly.
When switching to MPV mode all alarms, except from High/Low Pres-
sure Alarms, are automatically switched off to avoid false alarm trigger-
ing (when changing between profiles the alarm settings does not change
to Off, but stays as defined in the profiles)
CPAP – Continuous Positive Airway Pressure
In CPAP mode the Vivo 65 is applying a continuous positive pressure to
the airways. The flow will automatically be adjusted to maintain the set
CPAP level. CPAP mode can only be used when a leakage circuit is
selected.
Sigh
When the Sigh feature is enabled the Vivo 65 will deliver a Sigh breath as
configured by the prescriber at a certain preset frequency. A Sigh is a
breath where an increased % of the set pressure or volume is delivered to
the patient.
In Volume modes the Vivo 65 can deliver a Sigh using 125%, 150%,
175% or 200% of the set Tidal Volume. The Tidal Volume that will be
used during the Sigh for the selected % will be displayed on the Setting
screen.
In Pressure modes the Vivo 65 can deliver a Sigh using 125%, 150%,
175% or 200% of the set Inspiratory Pressure. The Inspiratory Pressure
that will be used during the Sigh for the selected % will be displayed on
the Setting screen.
• The alarm levels must be assessed and adjusted considering the
patients condition and treatment settings.
• MPV breath mode shall be used with a mouthpiece interface only.
• MPV breath mode shall not be used with ventilator dependent
patients.
Содержание 65
Страница 202: ...202 Accessories Vivo 65 users manual Doc 006088 En US U 1b 15 14 19 21 22 24 25 18 20 27 13 17 16 26 28 29 12 23 ...
Страница 208: ...208 Patient Settings Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 210: ...210 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 211: ...211 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 228: ...228 Index Vivo 65 users manual Doc 006088 En US U 1b W Warning icon 6 Water trap 17 Weight Vivo 191 ...