101
How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
• Incorrect CO
2
zeroing will result in false gas readings.
• Replace the airway adapter if rainout/condensation occurs inside the
airway adapter.
• Only use airway adapters distributed by Breas.
• Do not apply tension to the CO
2
sensor cable.
• To keep secretions and moisture from pooling on the windows,
always position the CO
2
sensor in a vertical position with the green
LED pointing upwards.
• If an intentional leakage port is used, make sure that the CO
2
sensor
is placed between the patient interface and the leakage port.
• If a patient interface with integrated leakage is used, the monitored
CO
2
values may be influenced.
• The CO
2
sensor should be placed as close to the patient interface as
possible. However, a HME (Heat and Moisture Exchanger) should
be placed between the patient interface and the CO
2
sensor. This will
protect the airway adapter from secretions and effects of water
vapour and eliminates the need of changing the airway adapter.
The CO
2
monitoring automatically compensates for changes in ambient
barometric pressure.
Содержание 65
Страница 202: ...202 Accessories Vivo 65 users manual Doc 006088 En US U 1b 15 14 19 21 22 24 25 18 20 27 13 17 16 26 28 29 12 23 ...
Страница 208: ...208 Patient Settings Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 210: ...210 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 211: ...211 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 228: ...228 Index Vivo 65 users manual Doc 006088 En US U 1b W Warning icon 6 Water trap 17 Weight Vivo 191 ...