152 Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
CO
2
Sensor Failure/Disconnection Alarm (CO
2
Discon-
nected)
I
TEM
D
ESCRIPTION
Definition
A CO
2
Sensor Failure/Disconnection alarm
will be given when communication between
the Vivo 65 and the CO
2
sensor has been lost.
Check the CO
2
sensor.
Priority
High
Possible cause
• CO
2
Sensor disconnected.
• Failure in the CO
2
sensor.
Ventilator action
The Vivo 65 will continue treatment with the
same settings.
Indication
The alarm is given audibly with a tone and
visibly by the red alarm LED and a display
message.