109
Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Audible Signal Pause and Reactivation
Alarm Reset
An alarm will automatically be reset once the cause of the alarm has been
corrected.
6.2 Operator’s Position
For receiving the audible part of an alarm, the operator’s position should
be within audible range from the Vivo 65, depending on the set audible
alarm level.
For receiving the visual part of an alarm and its priority, the operator’s
position should be within a distance of 13 feet (4 meters) from the
Vivo 65, and within an angle of 30° to the normal of the Vivo 65 display.
The audible signal can be paused
for 60 seconds by pressing the
Audio Pause button. The audible
signal can be reactivated by press-
ing the Audio Pause button again.
If a new alarm condition occurs
during the audio pause period, the
audible signal will be reactivated.
If an alarm condition cannot be corrected, discontinue use and refer the
Vivo 65 for service.