67
How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
Inspiratory Trigger (Insp. Trigger)
I
TEM
D
ESCRIPTION
Definition
The inspiratory trigger setting defines the
patient’s effort required to initiate a ventila-
tor assisted breath. When the patient starts a
breath, an increasing flow is created in the
patient circuit. If the patient’s effort reaches
the set inspiratory trigger level an inspiration
is initiated. If the patient cannot trigger a
breath, the ventilator will deliver breaths
according to the set Backup Rate, Breath
Rate or SIMV Rate.
The Assisted breath modes in PCV and
VCV are turned off if the inspiratory trigger
is set to Off.
Modes
PSV, PSV(TgV), PCV(A), PCV(A+TgV),
PCV-SIMV, PCV-MPV, VCV(A), VCV-
SIMV, VCV-MPV
Setting min
1
Setting max
9 (PSV, PCV-SIMV, PCV-MPV, VCV-SIMV,
VCV-MPV),
9, Off (PCV & VCV)
Setting resolution
1
(setting 1 is the most sensitive and 9 is the
least sensitive)