112 Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Low Pressure Alarm
I
TEM
D
ESCRIPTION
Definition
A Low Pressure alarm will be given when the
Vivo 65 pressure fails to reach the low pressure
alarm limit for 15 ±0.5 seconds.
In MPV breath mode the alarm will be given
when the pressure fails to reach the limit during
inspiration.
Priority
High
Possible cause
• Disconnection of patient circuit.
• Mismatch between Inspiratory Pressure/
CPAP and alarm setting.
• Leakage from the mask or other components
of the patient circuit.
Setting min
1 cmH
2
O
Setting max
50 cmH
2
O (Adult), 35 cmH
2
O (Pediatric)
Setting resolution
0.5 below 10 cmH
2
O, 1.0 above 10 cmH
2
O
The Low Pressure alarm setting is also visible
with a red line in the pressure bar:
Ventilator action
The Vivo 65 will continue treatment with the
same settings.
Indication
The alarm is given audibly with a tone and visibly
by the red alarm LED and a display message.