144 Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Circuit Type/Insert Mismatch Alarm
I
TEM
D
ESCRIPTION
Definition
A Circuit Type/Insert Mismatch alarm will be
given when the Vivo 65 detects an incorrect
combination of the selected patient circuit
type setting and the insert attached to the ven-
tilator.
Priority
Medium
Possible cause
The Vivo 65 is set to one of the following
incorrect combinations:
• Patient circuit type: Dual limb circuit / Sin-
gle limb insert
• Patient circuit type: Exhalation valve (sin-
gle limb) circuit / Dual limb insert
Ventilator action
The Vivo 65 will continue the treatment with
the same settings.
Indication
The alarm is given audibly with a tone and
visible by the yellow alarm LED and a display
message.