How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
CO
2
Zeroing
CO
2
zeroing is recommended when changing the airway adapter. Besides
from that, zeroing only needs to be performed when an offset in moni-
tored CO
2
values is observed, or when a CO
2
sensor accuracy unspeci-
fied message is displayed.
Maintenance
No periodical maintenance is required for the CO
2
sensor.
5
Make sure to position the CO
2
sensor with the LED pointing
upwards.
The CO
2
sensor is not intended to be in contact with the patient body.
CO
2
zeroing can be performed from the “FiO
2
/CO
2
Calibration” page under the
“Others” section.
LED S
TATUS
D
ESCRIPTION
Steady green light
System OK
Flashing green light
Zeroing in progress
Steady red light
Sensor error
Flashing red light
Check adapter