159
Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Humidity Compensation Lost Alarm (Humidity Comp Lost)
I
TEM
D
ESCRIPTION
Definition
An Humidity Compensation Lost alarm will
be given when the automatic humidity com-
pensation is out of order.
Priority
Medium
Ventilator action
The Vivo 65 will continue treatment with the
same settings. The accuracy of the volume
measurement may be impaired.
Indication
The alarm is given audibly with a tone and
visibly by the yellow alarm LED and a display
message.
Содержание 65
Страница 202: ...202 Accessories Vivo 65 users manual Doc 006088 En US U 1b 15 14 19 21 22 24 25 18 20 27 13 17 16 26 28 29 12 23 ...
Страница 208: ...208 Patient Settings Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 210: ...210 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 211: ...211 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 228: ...228 Index Vivo 65 users manual Doc 006088 En US U 1b W Warning icon 6 Water trap 17 Weight Vivo 191 ...