PART 2 - User manual
22
DAILY MAINTENANCE
VENDLET Standard is a reusable medical device intended for
long-term use, requiring safe procedures in connection with
cleaning and maintenance.
Daily inspections
VENDLET Standard must be checked daily in accordance with
the checklist below. If a ”Yes” checkbox is marked, this indi-
cates a desired result. If a ”No” checkbox is marked, this indi-
cates a risk in which case VENDLET Standard must not be used.
In case of malfunctions, faults, missing parts, or improper
fitting, contact a person with relevant training for inspection or
repair.
Checklist for daily inspections
In case of No, do
not use VENDLET
Standard
1
Is VENDLET Standard free from visible damage and crooked brackets and fittings?
Yes
No
2
Do the bars raise and lower as they should?
Yes
No
3
Are the nylon straps attached to the bed frame and all the support bars?
Yes
No
4
Is the turning sheet in good condition and does the sheet run over the bars?
Yes
No
5
Is the slide sheet fastened around the the edge of the mattress support platform and in good condition? Is the slippery side
facing upwards?
Yes
No
6
Double-click the
on the hand control. Does the green diode on the hand control illuminate?
Does everything work when you press the matching hand control buttons?
Yes
No
7
Has it been verified that the condition of the client and the mattress/bed/client combination have not changed relative to
when VENDLET Standard was mounted?
Yes
No
8
Are the caregivers familiar with the instructions of this manual and have they received training in the use of VENDLET Stan-
dard?
Yes
No
9
Is the quick guide located at an easily accessible spot near the bed?
Yes
No
10
Has it been verified that the date of the next annual inspection has not been exceeded?
Yes
No