PART 3 - Technical manual
37
Battery backup
Recharging during storage
First recharge of the battery must be no later than 12 months
after production date stated on the label. Hereafter the battery
must be recharged at least every 12 months.
DO NOT
•
Heat or burn the battery
•
Short circuit the battery
•
Expose the battery to high impact
•
Crush or puncture the battery
•
Use batteries with signs of damage or corrosion
•
Change or store the battery near combustible material
•
Expose the battery to water or other liquids when un-
plugged
LED functionality
LED
Indication of operating
Solid yellow
Charging
No LED Light
Fully charged
Flashing yellow
Error during charging
Buzzer functionality
The buzzer will make a warning when a button on the hand
control is pressed and the battery capacity is low.
Power consumption
VENDLET Standard
Standby
5 Watt
Tighten sheet (without person)
60 Watt
Slacken sheet
15 Watt
Moving a person at 100 kg
75 Watt
Moving a person at 250 kg
160 Watt
Raise bars
110 Watt
Lower bars
65 Watt
TECHNICAL DATA
In accordance with the Medical Device Regulation (EU)
2017/745, VENDLET Standard is a class I product.
VENDLET Standard
Product no.
6300000
Max. user weight
250 kg
Control unit:
Power supply
SMPS 350W
Mains supply
100-220 Volt AC @ 60 Hz or
100-230 Volt AC @ 50 Hz
Power rating
Max 350 W
Protection
Class 1
IP class
IPX6 Washable
Intermittence
10%, max 2 min / 18 min.
Output voltage
24 Volt DC
Weight:
Total system
approx. 42 kg
Red side
approx. 20 kg
Blue side
approx. 20 kg
Control unit (incl extension)
approx. 2 kg
Battery (accessory)
approx. 0,7 kg
Length:
200 - 220 cm (telescopic)
Width:
May be adapted to beds between
70 and 130 cm
Expected service life
10 years
Noise level max. load
55 db(A)
Turning sheet:
Regular width and length
270 cm / 183 cm
Material
100 % cotton
Slide sheet:
Regular width and length
95 cm / 210 cm
Material
100% nylon
System service check
Annually
Battery (optional):
Li-lon 2,25 Ah / 24 Volt DC
Duty Cycle:
5 %,
1 minute continuous use followed by
19 minutes not in use
Charging:
With integrated charger in battery
Charging time:
Approx. 10 hours
Operating temperature:
+ 5 °C to + 30 °C
Storage temperature:
- 10 °C to + 40 °C (+ 10 °C to + 25 °C
recommended) The batteries must be
stored in an applicable storage room
without direct sunlight
Relative humidity:
20 % RH to 80 % RH, non-condensing
Atmospheric pressure:
700 to 1060 hPa
IP class
IPX6 Washable
Содержание STANDARD