VEMCO - VR2C Manual
9
Jul 2018
51
5.2.2.2
Disconnecting and Removing Battery
After opening the VR2C case as described in “Opening the VR2C Case” (section 5.2.2.1), follow the steps
listed below to disconnect and remove the battery.
Place the two sections of the VR2C on a stable,
level surface. Be careful not to separate the upper
and lower units
too much or
undue strain
will be placed
on the internal
wiring harness.
STEP 1
Disconnect the 2-pin
battery connector by
pressing the release
lever on the connectors
while pulling them apart.
A small flat screw driver
may be helpful with the
release lever.
STEP 2
Disconnect the 6-pin
connector by pressing the
release lever on the
connectors while pulling them
apart. Again, a small flat
screw driver may be helpful
with the release lever.
Set the upper housing aside
in a safe location.
STEP 3
Hold the lower housing
vertically so the underwater
connector is pointing down
(battery wires point up).
Press down on the exposed
end of the battery until the
battery is no longer touching
the O-ring.
STEP 4
Remove the O-ring from the housing and slide it
along the battery wires.
STEP 5
Tip the lower housing
and grasp the battery as
it slides out of the
housing.