Logging Options
3.1.4.1
Ancillary Logging
The VR2C has the ability to measure the temperature and/or tilt
of the receiver and store this data in the receiver’s memory at an
interval you specify. Keep in mind that the VR2C memory will
fill quicker when this additional information is selected to be
stored.
1.
Select “Logging Options”
2.
Select the desired information (temperature, tilt, or both).
3.
Enter how often the data is to be stored in memory in the “Logging Period” box.
3.1.4.2
Diagnostic Logging
Diagnostic Logging is an advanced setting feature which should
only be selected when directed by VEMCO. This setting will
log information about the receiver that may be helpful for
VEMCO support staff when assisting a customer during the
initial VR2C deployment setup.
The Diagnostic Logging settings window allows the user to
specify both how often and for what duration the diagnostic
information will be stored to memory.
Two
baud rates are shown in the information given on the VR2C Receiver Tab page. The “Current
Baud” value is the baud rate used to communicate with VUE during this session. The “RTM Port”
information includes the VR2C permanent baud rate, which may not be the
same as the “Current
Baud” value.
Baud Rates
RTM baud rate for communication
with external host.
Current baud rate for
communication with VUE.
Only enable if directed to do so by VEMCO.